Proline Kryomats
104
YAUE0007 / 10/2/2018
Command
Explanation
OUT_MODE_00_X
Keyboard Master: 0 = free / 1 = locked (corre
sponds to “KEY”).
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external Analogue / 3 = external
Serial.
OUT_MODE_03_X
Keyboard Command remote control: 0 = free / 1 = locked.
OUT_MODE_04_X
Setpoint offset source: 0 = normal / 1 = ext. Pt / 2 = ext. analog / 3 = ext. serial.
START
Switches the unit on (after Standby).
See safety information (
STOP
Switches the device into Standby (pump, heater, cooling unit OFF).
RMP_SELECT_X
Selection of the program (1
– 5) to which the further instructions apply. When the
unit is switched on, program 5 is selected automatically.
RMP_START
Start the programmer.
RMP_PAUSE
Hold (pause) the programmer.
RMP_CONT
Restart the programmer after pause.
RMP_STOP
Terminate the program.
RMP_RESET
Delete the program (all Segments).
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_
X
Set a programmer segment (temperature, time, tolerance and pump level). A
segment is added and appropriate values are applied to it.
RMP_OUT_02_XXX
Number of times the program runs: 0 = unlimited / 1
– 250.
For ”_“ use also ” ” (blank character).
R
esponse from thermostat ”OK“ or in case of error ” ERR_X“ (RS-485 interface
e.g. “A015_OK” or in case of error ”A015_ERR_X”.).
The command from the computer must be terminated with CR, CRLF, or LFCR.
The response of the thermostat is always terminated with CRLF.
After each command sent to the thermostat, it is necessary to wait for the reply
before sending another command. This ensures that the sequencing of inquiries
and answers is clear.
CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
Permitted data formats:
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
8.3.6
Read commands (Data requested from the thermostat)
Command
Explanation
IN_PV_00
Read bath temperature (outflow temperature).
IN_PV_01
Indication of the controlled temperature (int./ ext. Pt/ ext. Analogue/ ext. Serial).
IN_PV_03
Read external temperature TE (Pt100).
IN_PV_04
Read external temperature TE (Analogue input).
IN_PV_05
Read bath level.
IN_PV_10
Read bath temperature (outflow temperature)
in 0.001 °C.
IN_PV_13
Read external temperature TE (Pt100)
in 0.001 °C.
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......