Proline Kryomats
62
Starting up
YAUE0007 / 10/2/2018
7.6.2
Defining temperature limits
With this function, it is possible to define a minimum and a maximum temperature in which the
thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input
can be prevented which may damage the heat transfer liquid or the apparatus. For example, if water
were used as the heat transfer liquid, +95 °C would be practicable as the maximum temperature and
+5 °C as the minimum temperature.
Command
Limits
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Limits
.
The minimum and maximum
temperatures are displayed.
T il (min)
is currently active.
Select the limit to be changed with
or
and confirm with
.
Enter the desired limit temperature.
Accept the change with
,
or quit the window with
without making
changes.
Pump Menu
End
T
set
T
fix
T il
(min)
-50.0 °C
T ih
(max)
90.0 °C
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Lower limit (T il)
-50.
0
Min: -100.0 °C Max: 202.0 °C
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......