The following information refers to the Ethernet
‑
as well
as the RS 232/485
‑
interfaces.
Command
Meaning
OUT_PV_05_XXX.XX
Specifying external temperature via interface
OUT_SP_00_XXX.XX
Setpoint transfer with max. 3 digits before the decimal point and max. 2
digits afterwards
OUT_SP_02_XXX
State cooling (0 = OFF / 1 = ON / 2 = AUTOMATIC)
OUT_SP_04_XXX
Upper limit of TiH outlet temperature
OUT_SP_05_XXX
Lower limit of TiL outlet temperature
OUT_PAR_00_XX.X
Setting control parameter Xp
OUT_PAR_01_XXX
Setting control parameter Tn (5–180 s; 181 = Off)
OUT_PAR_02_XXX
Setting control parameter Tv
OUT_PAR_03_XX.X
Setting control parameter Td
OUT_PAR_04_XX.XX
Setting control parameter KpE
OUT_PAR_05_XXXX
Setting control parameter TnE (0–9000 s; 9001 = Off)
OUT_PAR_06_XXXX
Setting control parameter TvE (5 = OFF)
OUT_PAR_07_XXXX.X
Setting control parameter TdE
OUT_PAR_09_XXX.X
Setting the correction value limit
OUT_PAR_10_XX.X
Setting control parameter XpF
OUT_PAR_14_XXX.X
Setting the setpoint offset
OUT_PAR_15_XXX
Setting control parameter PropE
OUT_MODE_00_X
Keyboard: 0 = free / 1 = blocked (corresponds to: "KEY")
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external analogue / 3 =
external serial
OUT_MODE_03_X
Keyboard remote control unit Command: 0 = free / 1 = blocked
OUT_MODE_04_X
Setpoint offset source: 0 = normal / 1 = external Pt / 2 = external ana-
logue / 3 = external serial
START
Switches on device (from standby)
STOP
Switches machine to standby (pump, heater, cooling unit off)
RMP_SELECT_X
Selection of program (1–5) that other commands are to refer to. After
switching on the device, program 5 is selected.
Before starting up
Variocool
35