Error
Description
ERR_2
Incorrect input (e.g. buffer overflow)
ERR_3
Incorrect command
ERR_5
Syntax error in the value
ERR_6
Impermissible value
ERR_8
Module or value not available
ERR_30
All segments of the programmer are occu-
pied
ERR_31
Setpoint input not possible, analogue set-
point input is set to ON.
ERR_33
External temperature probe not present
ERR_34
Analogue value not available
ERR_35
Automatic selected
ERR_36
Setpoint input not possible, the programmer
is running or on hold.
ERR_37
Starting the programmer not possible, ana-
logue setpoint input selected.
4.4.8 Cable and test of interface RS 232
Computer
Thermostat
Signal
9-pin Sub-D female con-
nector
25-pin Sub-D female
connector
9-pin Sub-D female con-
nector
Signal
with hard-
ware hand-
shake
without
hardware
handshake
with hard-
ware hand-
shake
without
hardware
handshake
with hard-
ware hand-
shake
without
hardware
handshake
RxD
2
2
3
3
2
2
TxD
TxD
3
3
2
2
3
3
RxD
DTR
4
20
4
DSR
Signal
Ground
5
5
7
7
5
5
Signal
Ground
DSR
6
6
6
DTR
RTS
7
4
7
CTS
CTS
8
5
8
RTS
With hardware handshake: Use a straight-through cable and not a
null modem cable when connecting a thermostat to the PC. The
RS 232 interface can be connected directly with the PC using a
straight-through cable.
Without hardware handshake: Set the corresponding operating
mode on the PC.