Fig. 7: Flange connector (front) in idle state
1
Normally open contact
2
Centre
3
Normally closed contact
Front view of the flange connector or of the coupling socket on the
solder side
Idle state
n
The alarm output is in resting state:
When the device is switched off,
after switching on if a fault is already evident at that time
(e.g. level too low)
and during normal operations when a fault occurs.
n
Pins 1 and 2 are open.
n
Pins 3 and 2 are closed.
GO state
n
The alarm output is in a sound state when no faults occur.
n
Pins 1 and 2 are closed.
n
Pins 3 and 2 are open.
Information on the settings for the alarm output can be
found in
Chapter 6.12.7 ‘Configuring alarm output’
Note the following:
n
The equipment connected to the low voltage inputs and low
voltage outputs must have be safely separated from dangerous
voltages according to DIN EN 61140. For example, using
double or reinforced insulation according to DIN EN 60730-1 or
DIN 60950-1.
n
Only use shielded connection cables. Connect shield to con-
nector case. Cover unused plug connections with protective
caps.
4.4.2 Installing the driver for the standard USB interface
Before starting up
Variocool
29