Code
English output
Description
15
Digital Input
Fault at the digital input / switch contact
20
T ext Ethernet
No signal for actual value via the Ethernet module
8.3 Low Level alarm
n
If the liquid level falls below the minimum level, an alarm
sounds.
n
Low Level
is shown on the screen. The device components
such as the pump are switched off by the electronics.
Personnel:
n
Operating personnel
1.
Refill lacking heat transfer liquid.
2.
Unlock the display with the OK button.
The device restarts.
8.4 Overtemperature alarm
This alarm can only occur for devices with heater.
n
The message
3 Overtemperature
is shown on the screen if the
temperature monitor trips.
n
The electronics switch off the components of the device.
n
The device outputs a two-tone signal.
Fig. 74: Alarm Low Level
Rectifying fault
Fig. 75: Overtemperature alarm
Faults
Variocool
85