Faults / Causes / Correction
Part 1 • Page 15-3
Part 1 – Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 1 • Ch
apter 15
15.2.3 Fault code
Alarm/Error
Cause / Condition
Characteristics
Internal RAM and CPU test
DI: for error:
--
Delay:
No
Self acknowledging:
No
Reset+Rem contr. ackn: No
Initiating operating phase
System
OFF
In case of a watchdog the system goes to reset
DI: for error:
--
Delay:
No
Self acknowledging:
No
Reset+Rem contr. ackn: No
Initiating operating phase
System
OFF
EEPROM defect
or
Data transmission to EEPROM interrupted
DI: for error:
--
Delay:
No
Self acknowledging:
No
Reset+Rem contr. ackn: No
Initiating operating phase
System
OFF
Real time clock defect
or
Data transmission to real time clock interrupted
DI: for error:
--
Delay:
No
Self acknowledging:
--
Reset+Rem contr. ackn: No
Initiating operating phase
System continues to run, no automatic operation possible
Motor protection switch PKZM “
16Q2”
pump M1
failed
DI: for error:
DI 8 = 0
Delay:
No
Self acknowledging:
--
Reset+Rem contr. ackn: --
Operating phase
Initiating operating phase/action
EcoRO Dia I (HT)
Dialysis mode:
RO continues to run with M2, after
-
After switchover by means of remote control, change from system off to dialysis mode
-
Or switchover by means of operating modes switch from 0 to 1, change from system off
to dialysis mode
-
Or switching off and on again by means of main switch and operating modes switch to 1,
start dialysis mode again
Standby mode:
RO continues to run with M2
Disinfection:
RO continues to run with M2
Cleaning:
RO continues to run with M2
(Eco)RO Dia II (HT)
Dialysis mode:
Initiate emergency mode RO II
Standard RO:
Message switchover K1 and press F2
RO with HotRO:
Message switchover K1 and press F2
Emergency mode RO I:
No function
Emergency mode RO II:
After switchover by means of remote control, change from emergency mode RO II to
system off
After switchover by means of operating modes switch from 0 to 1, change from system
off to emergency mode RO II
Switching off and on again by means of main switch and operating modes switch to 1,
start emergency mode RO II
Standby mode:
System OFF
Disinfection:
RO continues to run with the other pumps
Cleaning:
RO continues to run with the other pumps
HotRO:
Pump off, if Hot Rinse Alarm then standby mode break
wd. dd.mm.yy ss:mm
Alarm 05
Failure pump M1
Menu
wd. dd.mm.yy ss:mm
Alarm 04
RTC
Menu
wd. dd.mm.yy ss:mm
Error 03
EEPROM
Menu
wd. dd.mm.yy ss:mm
Error 02
Watchdog LT/ext.sup.
Menu
wd. dd.mm.yy ss:mm
Error 01
CPU
Menu