Faults / Causes / Correction
Part 1 • Page 15-7
Part 1 – Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 1 • Ch
apter 15
Limit value at pressure switch PSAH3 exceeded
DI: for error:
DI 16 = 0
Delay:
3 sec
Self acknowledging:
--
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
System OFF
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
System OFF
The level has dropped below LSHAL1 during
standby mode, break or intermediate rinsing.
or
a difference FISCAL 1 => FISCAL 2 is determined
during intermediate rinsing
DI: for error:
DI 12 = 0
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
No function
Emergency mode RO I:
No function
Emergency mode RO II:
No function
Standby mode:
System OFF
Disinfection: No
function
Cleaning: No
function
HotRO I/ HotRO I+II
No function
All motor protection switches on pumps have
failed
DI: for error: DI 8/9/10/11 = 0
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
System OFF
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
System OFF
No pulses from F FISCAL2
DI: for error: DI 17 + 18 = 0
Delay:
60 sec
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
System OFF
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
No function
wd. dd.mm.yy ss:mm
Error 19
No perm. available
Menu
wd. dd.mm.yy ss:mm
Error 18
Pumps failed
Menu
wd. dd.mm.yy ss:mm
Error 17
Leakage alarm
Menu
wd. dd.mm.yy ss:mm
Error 16
PSAH3 press. ROII
Menu Reset