Faults / Causes / Correction
Part 1 • Page 15-11
Part 1 – Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 1 • Ch
apter 15
Illogical switch state of the float switches LSHAL1
(reports tank empty) + LSHL2 (reports tank full)
DI: for error:
DI = 12 = 0
DI = 13 = 1
Delay:
5 sec.
Self acknowledging:
yes
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
No function
Standby mode:
System off, when level LSHAL1 is not reached, Error 17 leakage alarm
Disinfection: System
OFF
Cleaning: System
OFF
Hot RO I/ Hot RO I+II
System OFF
Wire break sensor or temperature 0°C
If Error 31 is active
(must be active for all HT versions)
DI: for error:
NTC
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
No function
Emergency mode RO I:
No function
Emergency mode RO II:
No function
Standby mode:
No function
Disinfection: No
function
Cleaning: No
function
Hot RO I/ Hot RO I+II
Cancel HotRO and change to cool-down phase
TISAH1: Temperature > 95°C
DI: for error:
CH5
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
No function
Emergency mode RO I:
No function
Emergency mode RO II:
No function
Standby mode:
No function
Disinfection: No
function
Cleaning: No
function
HotRO I/ HotRO I+II
Cancel HotRO and change to cool-down phase
TISAH2: Temperature > 95°C
DI: for error:
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
No function
Emergency mode RO I:
No function
Emergency mode RO II:
No function
Standby mode:
No function
Disinfection: No
function
Cleaning: No
function
HotRO I/ HotRO I+II
Cancel HotRO and change to cool-down phase
wd. dd.mm.yy ss:mm
Error 33
TISAH2 temp. too high
Menu
wd. dd.mm.yy ss:mm
Error 32
TISAH1 temp. too high
Menu
wd. dd.mm.yy ss:mm
Error 31
TISAH2 defect
Menu
wd. dd.mm.yy ss:mm
Alarm 30
Level detection VT