Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 15-12
Faults / Causes / Correction
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chap
ter 15
Max. heat-up time for RO I+II and/or RO II of
300 minutes has been exceeded
DI: for error:
--
Delay:
No
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
No function
Emergency mode RO I:
No function
Emergency mode RO II:
No function
Standby mode:
No function
Disinfection: No
function
Cleaning: No
function
HotRO I/ HotRO I+II
Cancel HotRO and change to cool-down phase
Alarm from external water monitor
DI: for error:
DI 22 = 0
Delay:
10 sec
Self acknowledging:
yes
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
System OFF
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I+II:
System off, after self-acknowledging proceed to HotRO I+II for cool down
(HotRinse SMART after water monitor alarm in Emergency Stop)
HotRO I:
System off, after self-acknowledging proceed to HotRO I for cool down
(HotRinse SMART after water monitor alarm in Emergency Stop)
wd. dd.mm.yy ss:mm
Alarm 35
Aqua Control
Menu
wd. dd.mm.yy ss:mm
Error 34
Max heating time
Menu