Maintenance and operating logs
Part 2 • Page 8-7
Part 2 – Supplementary Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 2 • Chapt
er 8
8.5
CLEANING LOG (reference to document E07FB18)
Execution:
1.
Instruct the person responsible about cleaning
2.
Disconnect the dialysis machine
3.
Clearly mark that the system is being cleaned
4.
Check Hydrowatch: red indicator is not visible
5.
Carry out cleaning using program mode “C”
6.
Note times
Cleaning agent used:
•
I have made sure that the permeate conductivity determined after cleaning is less than or equal to the permeate conductivity determined before
cleaning
•
The pH value in the concentrate has the same value before and after cleaning (±0.1 pH)
End of cleaning:
.........................
Signature customer: ...............................................
Place/date
.........................
Signature technician:..............................................
Customer
Street
Postal code and city
Cleaning ordered by
on
Cleaning initiated by:
on
IMPORTANT The warning and safety instructions of the operating manual must always be followed!
The following reverse osmosis system was cleaned
SN.:
Before cleaning
After cleaning
Unit
Permeate conductivity v
µS/cm (25°C nLF)
pH value in the concentrate
--
Permeate hourly throughput
l/h
Permeate temperature
°C
Expiry date
Quantity
Concentration
Circulating time
Reaction time
Flushing time
NOTICE
It has been ensured that disinfection will be carried out after cleaning of the reverse osmosis
system