Turning the system on
Part 1 • Page 9-1
Part 1 – Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 1 • Chapt
er 9
9.
Turning the system on
Operating status and phase
To supply the system with power, insert the mains plug and turn the
main switch to ON.
Switch the operating mode switch to 0 = System off.
After an initial test of the control unit, the display will show the current operating
mode “System off”.
Operating mode switch, switch position 0:
Switch position 0 means the system is off, all fault messages are reset.
Operating mode switch, switch position 1:
Switch position 1 means dialysis mode.
The remote control input and menu 3 auto on/off are not taken into
account.
Operating mode switch, switch position 2:
Switch position 2 means standby mode.
The remote control and menu 3 auto on/off can be used to switch
between standby mode and dialysis mode, emergency mode RO I,
emergency mode RO II.
Fault message
(wd. dd.mm.yy hh:mm:ss = weekday, day, month, year / hour, minute).
If a fault is determined in this mode of operation, the fault message
(“Alarm” or “Error”) is given in plain text.
Depending on the fault type, it can be acknowledged by pressing the >Reset<
button.
wd dd.mm.yy ss:mm
EcoRO Dia II
CPU-test
Menu
wd dd.mm.yy ss:mm
System off
Menu
wd dd.mm.yy hh:mm
Alarm
Menu Reset
NOTICE
After the power supply has been interrupted, the system always switches to the mode of
operation that was last active.