Standby mode
Part 1 • Page 11-1
Part 1 – Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 1 • Ch
apter 1
1
11. Standby mode
Standby mode is started by switching the operating mode selection switch to position 2 or, if it is already on
2, by using a remote control or the auto on/off function to switch the system to standby mode.
When standby mode is activated, shut-off rinsing is carried out first. With activated IRS (menu 2), this
begins with impulse back washing of the 1st stage.
As a result, the risk of bacteria growth in the loop and on the membrane is minimized.
NOTICE
Dialysis is not possible during “standby mode”.
You can switch to dialysis mode at any time by pressing the “Dial” key (F4) or turning the
operating mode selection switch to “1”.
Operating phase shut-off rinsing / concentrate discharge
The tank is emptied after shut-off rinsing.
The tank is filled in 2 steps.
This is followed by a first intermediate rinsing.
Shut-off rinsing takes place within the following sequence of sub-proc-
esses: Shut-off rinsing – emptying tank – filling tank 1 – filling tank 2 –
intermediate rinsing.
Break operating phase
After completion of the shut-off rinsing, a night pause is taken between
the intermediate rinsing operations.
Intermediate rinsing (flushing) operating phase
The repeated intermediate rinsing operations are similar to dialysis
mode. Permeate is produced and circulated through the loop. Perme-
ate tapping is not allowed, however.
wd dd.mm.yy hh:mm
Standby mode
Shut-off rinsing
Menu D/C Dial
wd dd.mm.yy hh:mm
Standby mode
Filling tank 1
Menu D/C Dial
wd dd.mm.yy hh:mm
Standby mode
Filling tank 2
Menu D/C Dial
wd dd.mm.yy hh:mm
Standby mode
Break
Menu D/C Dial
wd dd.mm.yy hh:mm
Standby mode
Intermediate flush
Menu D/C Dial