Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 6-2
Functions
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chapter 6
6.1.2
EcoRO Dia I (HT)
Figure 6-2:
Process diagram EcoRO Dia I (HT)
Legend
VF:
Prefilter
TISHAH1: temperature feed water ROI
PH2:
sample valve loop return
SF:
Sterilefilter
TISAH2:
temperture concentrate ROI (HT)
PH3:
sample valve softened water
LSHAL1: Level switch low; pump protection
LSHL2:
Level switch
PI 1-6:
Manometer
K1:
butterfly valve emergency operation
PSAH1:
pressure too high loop
K4:
butterfly valve emptying tank
M1:
Pump 1
PISAL4:
pressure too low loop return
K5:
butterfly valve
M2:
Pump 2
K6:
butterfly valve to loop
CISAHH1: conductivity permeat
K7:
butterfly valve from loop
Y2:
Solenoid valve concentrate ROI
CIS1:
conductivity raw water
MM1.1-1.2: Membrane module
Y3:
Solenoid valve pressure generation (IRS)
FISCAL1: flow meter permeate to loop
VL:
storage tank
Y4:
Solenoid valve permeat ROI
FISCAL2: flow meter permeate loop return
Y5:
Solenoid valve permeate to loop
FISCAL3: flow meter concentrate to drain
Option:
Y6:
Solenoid valve rinsing RO I (IRS)
UV2:
overflow valve
Y9:
Solenoid valve concentrate to drain
UV1:
overflow valve loop
Y30:
solenoid valve share force rinsing
Y10:
Solenoid valve feed
RSV:
check valve
PH1:
sample valve
NV4:
adjusting valve concentrate RO I
DG:
surge tank
PI2-4:
with analog signal to PV