- 32 -
PROGRAM SELECTION: confirm
and use the arrow keys to select the desired program:
: basic program, setpoint management only.
Set the approval status:
-
: not approved program;
-
: approved program, single interval (Dir. 2014/31/EU, art. 1)*;
-
: approved program, multi-interval (Dir. 2014/31/EU, art. 1)*.
* Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section INSTRUMENT
COMMISSIONING).
Set the active regulation on the instrument (only if
have not been set):
-
: approved program according to OIML (see the standards observed in the regions listed
in the TECHNICAL SPECIFICATIONS section of the approved manuals);
-
: approved program according to NTEP (see the standards observed in the regions listed
in the TECHNICAL SPECIFICATIONS section of the approved manuals).
By confirming, the instrument is restored to default and data is erased.
If you do not have a specific manual for the newly set program, you can request it
to technical assistance.
KEYPAD OR DISPLAY LOCKING
Press
immediately followed by
and hold them down for about 5 seconds (this operation is also
possible via the MODBUS and ASCII protocols):
-
: no lock.
-
: keypad lock: if active, when a key is pressed the message
is displayed for 3 seconds.
-
: keypad and display lock: if active, the keypad is locked and the display shows the instrument
model (weight is not displayed); by pressing a key the display shows
for 3 seconds.