- 16 -
EQUALIZATION
At the end of the equalization you must perform the
TARE WEIGHT ZERO SETTING
and, if necessary,the
REAL CALIBRATION
.
REAL EQUALIZATION
Use a sample weight equal to
at least 50%
of the single load cell capacity.
: unload the scale, wait for stability and confirm by pressing ENTER.
: place the sample weight on load cell 1, wait for stability and confirm by pressing ENTER.
: place the sample weight on load cell 2, wait for stability and confirm by pressing ENTER.
Repeat the operation for each connected load cell.
If equalization is successfully completed, the display shows
, confirm by pressing ENTER to
exit; if an error occurs, the display shows
, confirm by pressing ENTER and repeat the
procedure.
THEORETICAL EQUALIZATION
÷
: set the sensitivity for each load cell, leaving it at 0 for non-active channels.
EQUALIZATION COEFFICIENTS
÷
: it displays the equalization coefficients calculated for each active channel.
EQUALIZATION DELETION
Confirmation is requested (
), press ENTER to reset the equalization or press ESC to cancel
the command.
+
+
+
+