- 25 -
Minimum and maximum values which can be set for zero and full scale corrections
:
ANALOG OUTPUT TYPE
Minimum
Maximum
0÷10 V
-0.150
10.200
0÷5 V
-0.150
5.500
0÷20 mA
-0.200
22.000
4÷20 mA
-0.200
22.000
NOTE:
the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero (
) may be greater than the weight set for the analog full scale
(
). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000
= 0 analog output 0÷10 V
Weight =
0 kg
analog output = 10 V
Weight = 5000 kg
analog output =
5 V
Weight =10000 kg analog output =
0 V
SERIAL COMMUNICATION SETTING
-
: it disables any type of communication (default).
-
: MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
: ASCII bidirectional protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
-
-
: continuous weight transmission protocol (see Communication protocols manual), at the
frequency set in
item (from 10 to 300).
-
(set:
=
,
=
).
-
(set:
=
,
=
).
-
: continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series remote
displays; the remote display shows the net weight or gross weight according to its settings (set:
=
,
=
,
=
).
-
: continuous weight transmission protocol to RIP675, RIP6125C series remote displays;
the remote display shows the net weight or gross weight according to its settings (set:
=
,
=
,
=
).
-
: continuous weight transmission protocol to RIP675, RIP6125C series remote displays
(set:
=
,
=
,
=
).
When the remote display is set to gross weight:
-
if the instrument displays the gross weight, the remote display shows the gross weight.
+