- 26 -
-
if the instrument shows the net weight, the remote display shows the net weight alternated
with the message
.
-
: printer.
-
: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
-
: instrument address (from 1 to 99; default: 1).
-
: maximum transmission frequency (10 20 30 40 50 60 70 80 100
200 300; default: 10); to be set when the
transmission protocol is selected.
Maximum setting frequency
(
)
:
-
20 Hz with minimum baud rate 2400 baud.
-
40 Hz with minimum baud rate 4800 baud.
-
80 Hz with minimum baud rate 9600 baud.
-
100 Hz with minimum baud rate 19200 baud.
-
200 Hz with minimum baud rate 38400 baud.
-
300 Hz with minimum baud rate 38400 baud.
-
: delay in milliseconds which elapses before the instrument replies (from 0 to 200
ms; default: 0).
-
:
-
: no parity (default).
-
: even parity.
-
: odd parity.
-
: stop bit (1 2; default: 1).
-
: number of copies(1 9; default: 1).
-
: number of blank lines between one printout and the next.
-
: printing of custom heading from PC (
; default:
).
-
: connected printer type:
-
-
-
For more information about protocols and methods of communication, request the
proper manual to technical assistance.