16
Error Information and Treatment
It provides the error diagnostics and prompting information of this equipment. Users can judge and deal with problems
according to the prompting information and the solutions given in the following form.
prompting information
meaning of the information
solutions
CCC CCC
The result of measurement is beyond the
range.
OFF OFF
System gives the prompt when the STOP key
is interrupted accidentally
Err 01
When the guard is set enabled, press START
key without closing it or open the guard
artificially while the wheel is in rotating
measurement. In either of these two cases,
the balancer gives the prompt
Close the guard, or turn off the
guard function option in setting
items. However, because the laws
and regulations of safety protection
in different countries are not
completely the same, we suggest
not turning off the guard function
option.
Err 02
Prompt is given and measurement is stopped
when rotating speed is too low to meet the
basic measurement needs,
Problems of the electrical motor
shaft or the transmission belts.
Check and adjust.
Too light load also results in this
phenomenon, so please adjust the
load weight.
Err 03
The measurement rotation is in wrong
direction. This usually will appear in the
three-phase motor control balancer due to
sequence errors
Adjust the sequence of the
three-phase power.
ERR 10
Gauge error
Turn off the machine, return the
gauge to position 0, and then restart
it. If the error still exists, calibrate
the gauge following “Calibration
programs”
ERR CAL
The machine is not calibrated.
Users calibrate the machine
following “Setting programs”
ERS CAL
Factory maintenance error.
Contact the manufacturer.