INSTALL SEAL PURGE TUBING
Cut the end of tubing
squarely, using a razor
blade or tube cutter, if
necessary. Install the ¼”
O.D. tubing by inserting the
tubing into the Seal Purge
push-to-connect fitting.
Push in until the tubing
seats completely. Connect
the other end of the tubing
to a 60 – 70 psi CDA or N2
supply.
THE SEAL PURGE
PRESSURE MUST BE
PRESENT FOR THE
SYSTEM TO OPERATE.
DO NOT
OVERPRESSURIZE
THE SYSTEM OR
INTERNAL DAMAGE
MAY OCCUR.
Maximum pressure
should never exceed 90psi.
A “Need CDA” error will
appear in the run mode
status screen on the 650
controller if this
requirement is not met.
To remove tubing, turn off
pressure source. With the
pressure at 0psi, press in
tubing and the
inner collar
at the same time then pull
back on the tubing.