3. Specifications 

 

 
 

Getter type: 

Alloy of Zr/V/Fe 
2 beds (350 and 200 Celsius) 

Gases purified: 

Ar/He/Ne/Xe/Kr. Nitrogen version available. 

Impurities removed: 

H

2

O,O

2

,CO,CO

2

,N

2

,THC, H

2

, CH

4

 (Heated) 

H

2

O,O

2

,CO,CO

2

,H

2

 (room temperature) 

Impurity level : 

<1 ppb for each impurity 

Flow: 

200 cc/min (nominal) 

Gas connections: 

1/8’’ compression or 
1/8’’ VCR  

Recommended operating 
pressure: 

100 PSIG (689 kPAG) 

Recommended minimum 
operating pressure: 

10 PSIG (28 kPAG) optional 1 PSIG (7 kPAG) 

Supply: 

115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz 

Power consumption: 

Maximum 200 watts 

Dimensions: 

12” (304.8) high, 4.12” (104 mm) deep, 4.25”’ (108 mm) wide 

Weight: 

5 lbs (2.26 kg) 

 

 

4. Cautions and installation information 

 

4.1. Gas line connection 

 
Be sure to connect the upcoming gas to the inlet on the top of the purifier. The outlet is 
on the bottom. Connecting the gas line incorrectly will decrease the capability of the 
purifier to generate pure gas. 
 
1/8” Truelok or 1/8” Face Seal fittings are used to interface your gas line. 
 
Stainless steel tubing is recommended to use with this purifier. 
 

4.2. Installation steps 

  
1.

 

After removing the purifier from the shipping container and protective wrapping, 

ensure that both end caps are tight. 

2.

 

Make sure that fittings in the gas line match those supplied with the purifier. 

Summary of Contents for LDP1000

Page 1: ...LDP1000 Gas Purifier for noble gases and nitrogen USER S MANUAL V1 0 Printed in Canada Copyright 2010...

Page 2: ...ent 1 Forwarning 2 2 Warranty maintenance and service policies 3 3 Specifications 6 4 Cautions and installation information 6 4 1 Gas line connection 6 4 2 Installation steps 6 5 Lifetime and LED indi...

Page 3: ...tek makes assumption that all operators have taken the time to read this information prior to install operate and troubleshoot the purifier If any error is suspected by the reader please contact LDete...

Page 4: ...r the loss or damage directly or indirectly arising from the use of the equipment of goods from breach of any warranty or from any other cause All other warranties expressed or implied are hereby excl...

Page 5: ...VERED BY UMBRELLA INSURANCE AND OR INDEMNITY COVERAGE PROVIDED TO BUYER ITS ASSIGNS AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HERE UNDER Major force Seller is not liable for failure to perform...

Page 6: ...tions may apply to international shipments PROPRIETARY RIGHTS Buyer agrees that any LDetek s software firmware and hardware products ordered or included in the goods ordered are proprietary of LDetek...

Page 7: ...tion Maximum 200 watts Dimensions 12 304 8 high 4 12 104 mm deep 4 25 108 mm wide Weight 5 lbs 2 26 kg 4 Cautions and installation information 4 1 Gas line connection Be sure to connect the upcoming g...

Page 8: ...make sure you have the right operating voltage 7 With the power on the purifier let the gas vent to the atmosphere for about 3 hours 5 Lifetime and LED indication The LDP1000 getter has a lifetime of...

Page 9: ...nual 8 6 Drawings Wiring Diagram Cable 4 Cable 5 Cable 2 Cable 3 RS 232 Connector Cable 1 Thermocouple 1 Thermocouple 2 Yellow Led Green Led Red Led Heater 1 Heater 2 Pin A Corcom Pin C Module Pin B P...

Page 10: ...he lifetime is over Getter replacement procedure Step 1 Turn off the power on the LDP1000 unit Step 2 Remove the inlet and outlet tubing connections Step 3 Step 4 Remove the front part of the LDP1000...

Page 11: ...power up the unit for normal operation running NOTE MAKE SURE TO WELL PURGED THE LINE BEFORE TO CONNECT TO THE INLET 10 MINUTES PURGE DO NOT EXPOSE THE DEVICE TO AIR MAKE SURE TO QUICKLY CONNECT INLET...

Page 12: ...LDP1000 User s manual 11...

Reviews: