73
3.
If the green LED is OFF and power is correct. Check the fuses in the power inlet module. Be sure to remove
the power inlet voltage to avoid risks of electrical shock. Remove the fuse from the module to measure the
continuity of it
4.
Check the connection between the power supply and the panel PC. To do so, refer to drawing ‘’Top and
bottom rack section’’ of the drawing section 8.0. By accessing the front door, check the connections on the
panel PC
5.
If the connections are good, then it is required to measure the voltage on the supply connector on the Panel
PC. The voltage should be 19VDC
6.
Contact LDetek support (
support@ldetek.com)
with the results obtained in the previous steps
Defective 4-20mA
module
1.
Recalibrate the defective output according to the section 6.3.1.9
2.
Refer to drawing ‘’Motherboard PCB’’ of the drawing section 8.0 to identify the 4-20mA modules and
fuses. Check if the fuse is not blown. Remove the fuse to measure the continuity of it
3.
Swap the defective 4-20mA module with a known working module and test it
4.
Contact LDetek support (
support@ldetek.com)
with the results obtained in the previous steps
Unstable reading
1.
Make sure there is no active alarm. If there are alarms, they must be resolved
2.
Make sure the calibration has been done properly
3.
Provide the followings information to a LDetek support (
support@ldetek.com)
:
a)
S/N of the unit
b)
Certified gas calibration values of the span cylinder(s) and its balance gas
c)
Quality of the carrier gas used
d)
S/N of the gas purifier installed on the unit
e)
3 x screenshots of the 3 latest span gas chromatograms
f)
3 x screenshots of the 3 latest analysis on the sample gas
Summary of Contents for MultiDetek2
Page 1: ......
Page 2: ......
Page 3: ...MultiDetek 2 Trace impurities analyzer USER S MANUAL V3 9 Printed in Canada Copyright 2016 ...
Page 13: ...MULTIDETEK 2 User s manual 6 ...
Page 14: ...MULTIDETEK 2 User s manual 7 ...
Page 26: ...MULTIDETEK 2 User s manual 19 Figure 6 Typical Installation ...
Page 28: ...MULTIDETEK 2 User s manual 21 Figure 7 Same source actuation ...
Page 29: ...MULTIDETEK 2 User s manual 22 Figure 8 2 sources actuation ...
Page 31: ...MULTIDETEK 2 User s manual 24 Figure 9 O2 doping ...
Page 73: ...MULTIDETEK 2 User s manual 66 8 0 Backpanel terminal identification ...
Page 75: ...MULTIDETEK 2 User s manual 68 9 0 Drawings Schematics ...
Page 76: ...MULTIDETEK 2 User s manual 69 Unit dimensions and cut out dimensions ...
Page 82: ...MULTIDETEK 2 User s manual 75 MotherBoard PCB ...
Page 83: ...MULTIDETEK 2 User s manual 76 Motherboard PCB electrical connections ...
Page 87: ...74 ...
Page 88: ...75 ...
Page 89: ......
Page 90: ......