(3)
Setting C/N on/off ..............................."CO"
Function
Setting and querying C/N set in Step (4) on/off.
Syntax CO_n
(Setting,
reply)
CO_? (Query)
n C/N
On/Off
0 Off
1 On
(4)
Setting and querying C/N value.........."CN"
Function
Setting and querying C/N value added.
Syntax CN_i (Setting,
reply)
CN_? (Query)
i
C/N value: 0 to 31.0 (dB)
Effective down to one digit after the decimal point
(5)
Incrementing C/N value......................"CU"
Function
Incrementing C/N value in current steps.
Syntax CU (Setting)
(6)
Decrementing C/N value ...................."CD"
Function
Decrementing C/N value in current steps.
Syntax CD (Setting)
(7)
Setting preset steps for C/N value......"CL"
Function
Setting preset steps for incrementing and decrementing C/N value.
Syntax CL_n
(Setting,
reply)
CL_? (Query)
n
C/N Value Steps
0 0.1
1 1.0
2 10.0
(8)
Setting BER measurement................."BO"
Function
Setting and querying BER measurement.
Syntax BO_n
(Setting,
reply)
BO_? (Query)
n BER
Measurement
0 Off
1 On
*
Only when Source on the TS Source screen is set to PN, this program code is turned
on effectively and automatically.
8-9