2. SPECIFICATIONS
4
2.3.2
Waveform Output
Output Format
Range 1
Outputs half the input signal
Range 2
Active through output
Range 3
Active through output
Output Connectors
2 BNC connectors
Output Impedance
75
Ω
Maximum Output Voltage
±1.5 V (DC + peak AC, at 75
Ω
termination)
Output Voltage Accuracy
±2 % (DC)
Frequency Response
±5 % (DC to 5 MHz)
2.3.3
Control Connectors
PC Interface
Specifications USB
1.1
Port Type
B
Function
Transmits settings and measured data
Camera Control
Custom order feature
2.3.4
Automatic Self Check
Offset Calibration
Calibration Voltage
0 mV
Criterion
Range 1
±10 mV
Range 2
±5 mV
Range 3
±2.5 mV
DC Voltage Calibration
Calibration Voltage
Range 1
±2 V
Range 2
±1 V
Range 3
±0.5 V
Criterion
Range 1
±60 mV
Range 2
±30 mV
Range 3
±15 mV
AC Voltage Calibration (Level Calibration Only)
Calibration Voltage
Range 1
2 Vp-p
Range 2
2 Vp-p
Range 3
1 Vp-p
Frequency
3.6 MHz ± 0.5 %
Criterion
Range 1
±60 mV
Range 2
±60 mV
Range 3
±30 mV