14. STATUS DISPLAY (STATUS)
185
●
Setting the error detection upper limit
An error is detected when the level of the SDI input signal's luminance component exceeds
the specified value.
If you press F•D 1, the lower limit will be set to its default value of 109.4 % or 756.8 mV.
Procedure
STATUS (press and hold)
→
F•4 ERROR CONFIG
→
F•3 GAMUT ERROR
→
F•4 LUMA ERROR
→
F•2 UPPER LEVEL
→
F•D 1 VARIABLE
Settings When LUMA ERROR Is Set to %
Setting range: 90.8 to 109.4 % (The default setting is 109.4 %)
Settings When LUMA ERROR Is Set to mV
Setting range: 635.6 to 765.8 mV (The default setting is 765.8 mV)
●
Setting the error detection lower limit
An error is detected when the level of the SDI input signal's luminance component goes
below the specified value.
If you press F•D 1, the lower limit will be set to its default value of -7.2 % or -50.4 mV.
Procedure
STATUS (press and hold)
→
F•4 ERROR CONFIG
→
F•3 GAMUT ERROR
→
F•4 LUMA ERROR
→
F•3 LOWER LEVEL
→
F•D 1 VARIABLE
Settings When LUMA ERROR Is Set to %
Setting range: -7.2 to 6.1 % (The default setting is -7.2 %)
Settings When LUMA ERROR Is Set to mV
Setting range: -50.4 to 42.7 mV (The default setting is -50.4 mV)
14.6
Clearing Errors (ERROR CLEAR)
To clear errors, press F•5 ERROR CLEAR in the status menu.
STATUS (press and hold)
→
Figure14-13 Status menu
14.6.1 Clearing
Errors
To reset the error counts in the status display to zero, follow the procedure below.
Also, the LAPSED indication at the upper left of the screen resets to 00:00:00.
Procedure
STATUS (press and hold)
→
F•5 ERROR CLEAR