8. STATUS DISPLAY
164
4. The lip sync measurement screen is displayed.
Press STATUS
→
F•2
SDI ANALYSIS →
F•4 AV PHASE.
The time difference when the luminance level of the video signal (the G signal level when
the input signal is RGB) exceeds the specified value or when the audio level signal
exceeds the specified value is measured, and the results are displayed numerically and
graphically for each channel.
The measured value is displayed in units of time and frames. If the audio signal cannot
be detected, "UNLOCK" is displayed. If the audio signal cannot be measured correctly,
"MISSING" is displayed. Further, when the measured value is updated, an asterisk is
displayed next to the channel.
You can set the video signal measurement range, video signal luminance level, and
audio signal level using F•5 AV PHASE SETUP.
8.7.1
Selecting the Measurement Range
To select the graph measurement range, follow the procedure below.
Procedure
STATUS
→
F•2
SDI ANALYSIS →
F•4
AV PHASE →
F•1 SCALE MAX: 50ms / 100ms /
500ms / 1.0s / 2.5s
8.7.2
Updating the Measurement Screen
To update the measurement screen, follow the procedure below.
Procedure
STATUS
→
F•2 SDI ANALYSIS
→
F•4 AV PHASE
→
F•2 REFRESH