2. VIDEO SIGNAL WAVEFORM DISPLAY
27
2.6.4
Selecting the Y-Axis Measurement Unit
When F•2 XY SEL is set to Y, to select the Y-axis cursor measurement unit, follow the
procedure below.
Reference
COLOR MATRIX → section
2.7.1, “Selecting the Color Matrix.”
Procedure
WFM
→
F•5
CURSOR →
F•3 Y UNIT: mV / % / R% / DEC / HEX / HDR
Settings
mV:
The measurement unit is volts.
%:
The measurement unit is percentage.
When the composite display format is NTSC, 714 mV is 100 %. When the
composite display format is PAL, 700 mV is 100 %.
R%:
The amplitude will be measured as a percentage of the amplitude at the time
when you pressed F•5 REFSET.
DEC:
Values are displayed in decimal with 0 to 100 % expressed as 64 to 940.
This option cannot be selected when COLOR MATRIX is set to COMPOSIT.
C
B
C
R
signal measurement is not supported.
HEX:
Values are displayed in hexadecimal with 0 to 100 % expressed as 040 to 3AC.
This option cannot be selected when COLOR MATRIX is set to COMPOSIT.
C
B
C
R
signal measurement is not supported.
HDR:
Values are displayed as a percentage or cd/m
2
.
You can select this option when SER07 is installed and when HDR signals are
being measured.
For details, see section 6.1.2, “Cursor Display.”
2.6.5
Selecting the X-Axis Measurement Unit
When F•2 XY SEL is set to X, to select the X-axis cursor measurement unit, follow the
procedure below.
Procedure
WFM
→
F•5
CURSOR →
F•3 X UNIT: sec / Hz
Settings
sec:
The measurement unit is seconds.
Hz:
The measurement unit is frequency, with the length of one period set to the
distance between the two cursors.