5. PICTURE DISPLAY
79
10. Use function dial (F•D) to select the file to copy from the USB memory.
11. Press F•3 FILE LOAD
The user-defined correction table that you selected is copied from the USB memory to
USER_A. The copy operation is complete when the file list screen disappears and the
display returns to the measurement screen.
If a file has already been stored to USER_A, an overwrite confirmation prompt appears.
If you want to overwrite the current file, press F•1 OVER WR YES. Otherwise, press
F•3 OVER WR NO.
After you have copied a user-defined correction table, you can select it by pressing F•1
GAMMA SELECT on the CINELITE menu. A loaded correction table is displayed using the
name determined by its NAME keyword.
5.4.8
Displaying Synchronized Markers
To synchronize the markers on the vector display and video signal waveform display to
measurement points P1 to P3 and REF that you specify on the CINELITE display, follow the
procedure below. Synchronized markers can be displayed only when an f Stop screen
or %DISPLAY screen is shown in the same multi-screen display.
Markers cannot be displayed on the video signal waveform display under the following
conditions.
• When SWEEP is set to V or H SWEEP is set to 2H in the video signal waveform menu
• When COLOR MATRIX in the video signal waveform menu is COMPOSIT
Marker display will not work properly when waveforms are being displayed using an external
sync signal.
Procedure
PIC
→
F•4
CINELITE →
F•4 CINELITE ADVANCE: OFF / ON
CINELITE ADVANCE = ON
Figure 5-22 Displaying synchronized markers