background image

3

Thank you for purchasing this Lec Medical product. Before you first use this

Ward refrigerator please carry out the following actions:

Check that the refrigerator has not been damaged in any way during transportation.

If any damage is found it must be reported to our customer services department
immediately on 0844-815-3755

Ensure all packaging has been removed, including cardboard, polystyrene and any
tape used to hold shelves in place for transportation.

The refrigerator has been cleaned prior to despatch; however, we advise that it should
be cleaned using lukewarm water containing a mild detergent and a soft cloth.

We recommend that the refrigerator should be left in an upright position for 24 hours
prior to switching on.

This appliance is intended for professional use in applications such as hospital wards
etc.

This appliance is not intended for use by persons (including children) with reduced
physical, sensory or mental capabilities, or lack of experience and knowledge, unless
they have been given supervision or instruction concerning use of the appliance by a
person responsible for their safety.

Children should be supervised to ensure that they do not play with the appliance.

Positioning and safely using your Lec Medical refrigerator

Make sure the appliance is placed in a dry, well ventilated site, away from heat
sources.

The appliance must be placed on a level surface. If necessary, adjust the feet on the
product so the appliance remains level during operation.

The appliance should have sufficient clearance around it to provide adequate
ventilation – 10cm at each side, 6cm at the rear and 2.5cm at the top.

Before ANY maintenance activity, pull the plug out of the socket by gripping the
plug, DO NOT pull on the plug cord.

DO NOT use double-wire extension cords. If an extension cord is necessary make
sure it is a cord, which possess a protection safety certificate.

If the main power supply cord is damaged, it should be replaced by a qualified service
engineer.

DO NOT attempt any repairs to the appliance yourself (apart from those highlighted
in the Maintenance and Service Section). Repairs carried out by someone without the
relevant training are putting their personal safety at risk and will invalidate your
warranty.

Do not store explosive substances such as aerosol cans with a flammable propellant in
this appliance

WARNING:

this appliance is only to be operated by authorised personnel or

personnel who have correct knowledge on how the appliance operates.

Transporting and moving your Lec Medical refrigerator

The refrigerator must always be moved in the vertical position. The cabinet must not be tilted
anymore than 40°. In the event that the cabinet is tilted more than 40°, the power supply must
not be connected until the appliance has stood upright for at least 24 hours.

Summary of Contents for WSR151

Page 1: ...1 Ward Refrigerator Range Applicable models WSR288 WSR151 USER MANUAL...

Page 2: ...4 Product details 5 Controller overview 6 Viewing and resetting minimum and maximum temperatures 7 Changing the factory set point 8 Changing the temperature scale 8 Alarms 9 Maintenance and Servicing...

Page 3: ...l ventilated site away from heat sources The appliance must be placed on a level surface If necessary adjust the feet on the product so the appliance remains level during operation The appliance shoul...

Page 4: ...gerator and connecting power to the refrigerator Also make sure there is good ventilation in the room If you are in doubt please contact your supplier WARNING Keep ventilation openings in the applianc...

Page 5: ...Shelf 5 Wire Basket 6 Adjustable foot 2 at front of appliance There are also 2 roller castors at the rear of the appliance to ease product movement Additional or replacement parts are available from...

Page 6: ...6 Controller overview and operation 1 Menu 2 Lock keypad 3 Light not available on all models 4 Maximum temperature Up 5 Power 6 Mute 7 Minimum temperature Down 8 Set Enter 1 2 3 4 5 6 7 8...

Page 7: ...7 Viewing resetting minimum and maximum temperatures...

Page 8: ...e the set point 3 seconds followed by Once followed by 1 second current set point followed by Increase Decrease up to 10 times followed by 1 second Note set point can be increased or decreased in 0 1...

Page 9: ...ture goes outside the set parameters either too high or too low Should one of these alarms occur this can be muted by pressing on the controller The load temperature is a simulated temperature which i...

Page 10: ...WARNING Do not use any knives or sharp metallic objects to remove frost Alternatively a non scheduled defrost can sometimes be carried out as follows Cleaning Your refrigerator should be cleaned regu...

Page 11: ...as a wall 2 Remove the feet from the bottom right hand hinge figures 1 and 2 3 Remove the screws from the hinge figure 3 and remove hinge 4 Pull the door from the bottom to remove from the refrigerato...

Page 12: ...Remove the plastic cover figures 1 and 2 2 Unscrew and remove the light bulb figures 3 and 4 3 Replace with a similar 15W bulb 4 Replace the plastic cover Replacement light bulbs are available from o...

Page 13: ...APPLIANCE MUST BE EARTHED The flexible cord mains lead fitted to this appliance has three cores for use with a 3 pin 13 amp plug If a BS 1363 13 amp fused plug is used it should be fitted with a 13 am...

Page 14: ...ains insulation gases and refrigerant and must be disposed of in the appropriate manner Please contact the manufacturer or local authority for advice on how best to dispose of this product A relevant...

Page 15: ...k that the appliance is operating at an ambient temperature that meets the product specification on page 4 Air circulation at the back of the appliance is hampered 1 Move appliance further away from t...

Page 16: ...48 Marketing Enquiries email marketing gdpa co uk tel 0844 248 4661 Spares and After Sales Service email info gdha com tel 0844 815 3755 fax 0844 248 4123 Address Lec Medical Stoney Lane Prescot Merse...

Reviews: