SECTION 16:
RE-ISSUING LECKEY
PRODUCTS
Most Leckey products are assessed
and ordered to meet the needs of
an individual user. Before reissuing
a product we recommend that the
therapist prescribing the product
has carried out an equipment
compatibility check for the new
user and has ensured that the
product being re-issued contains no
modifications or special attachments.
A detailed technical inspection should
be carried out on the product prior to
re-issuing. This should be carried out
by a technically competent person
who has been trained in the use and
inspection of the product.
Please refer to section 15 for the
required checks to be carried out.
Before re-issuing a product, it is
recommended that it is disinfected
using one of the following procedures:
Note: Always clean the product first
to remove accumulated soiling prior
to disinfecting/sanitising.
1. Where possible, we recommend
the use of an automated “fogger”
disinfection system that is used
to completely cover the product
surfaces with a disinfectant
agent. The manufacturer’s
instructions for the chosen
system should be referred to prior
to use and followed exactly.
If in any doubt to the
continued safe use of
your Leckey product or
if any parts should fail,
please cease using the
product and contact our
Customer Care Team or
your local dealer as
soon as possible.
2. Where an automated disinfection
system is not available, all
product surfaces should be wiped
over with disinfectant surface
wipes or liquid disinfectant on a
soft cloth. Suitable products are
70% IPA, 70% Alcohol or diluted
bleach. The manufacturer's
instructions for the chosen
product should be followed.
Product upholstery and fabrics
should be cleaned thoroughly
in accordance with section,
Cleaning and Care Information, of
this manual.
Ensure a copy of the user manual
is supplied with the product. A copy
can be downloaded from our website
www.leckey.com
Leckey recommend that a written
record is maintained of all product
inspections carried out during the
reissue of the product.