ScanStation C10/C5, Troubleshooting
81
ScanStation C10/C5
Battery
Problem
Possible Cause(s)
Suggested Remedies
When switching on the
instrument or starting a
scan, the system switches
off automatically.
Capacity of battery is too
low.
Recharge or change
battery.
When switching on the
instrument or starting a
scan, the system switches
off automatically even
though it was totally
recharged.
Battery charger is defec-
tive.
Check the function of the
battery charger. Please
note the charging status
displayed on the battery
charger.
Damaged cable.
Examine the cabling and
pay attention to damages,
which for example can
cause loose contacts or
shirt circuits. Defective
circuits need to be
replaced. Only use supplied
power cables.