ScanStation C10/C5, Care and Transport
5.6
Adjustment of the Circular Level
On the instrument
step-by-step
1. Level up the instrument in advance with the electronic level, assuming that the
instrument is correctly adjusted. In the
Main Menu
go to
Status
,
Level and
Laser plummet
,
Level
to access the electronic bubble.
2. The bubble must be centered. If it extends beyond the circle, use an allen key to
center it with the adjustment screws. Turn the instrument slowly 200 gon (180°).
Repeat the adjustment procedure if the bubble does not stay centered.
After the adjustment, no screw shall be loose.
C10_026