ScanStation C10/C5, Care and Transport
93
On the tribrach
step-by-step
1. Level up the instrument with the electronic level, assuming that the instrument
is correctly adjusted. Remove the instrument from the tribrach. In the
Main
Menu
go to
Status
,
Level and Laser plummet
,
Level
to access the electronic
bubble.
2. The bubble of the tribrach must be centered. If it extends beyond the circle, use
the adjusting pin in conjunction with the two cross-headed adjustment screws to
centre it.
After the adjustment, no screw shall be loose.
C10_027