ScanStation C10/C5, Care and Transport
3. Press
Page
to access the
Laser Plummet
page. Switch on the laser plummet.
Inspection of the laser plummet should be carried out on a bright, smooth and
horizontal surface, like a sheet of paper.
4. Mark the centre of the red dot on the ground.
5. Turn the instrument through 360° slowly, carefully observing the movement of
the red laser dot.
6. If the centre of the laser dot describes a perceptible circular movement or moves
more than 3 mm away from the point which was first marked, an adjustment may
be required. Inform your nearest Leica Geosystems authorised service workshop.
Depending on brightness and surface, the diameter of the laser dot can vary. At
1.5 m it is about 2.5 mm.
The maximum diameter of the circular movement described by the centre
of the laser point should not exceed 3 mm at a distance of 1.5 m.