50
SPRINTER 100/100M/200/200M
Check & Adjust
Step 4
Aim at Staff A and press MEAS key. Measurement display, press ENTER key to
accept/confirm.
The new collimation error is displayed.
To accept the correction, press ENTER key.
If the reticle adjustment is required, reverse the staff B for the E scale bar reading then
adjust the reticle using the tool provided by following through the given instructions.
In the event where warning message e.g. "Invalid Measurement" is displayed for any
adjustment measurement, carry on measuring until a successful measurement is obtained
(Presumably instrument is setup correctly and external conditions are conducive) and
press ENTER key to accept the measurement.
MENU ADJ
ADJUST (4/4)
MEAS
MENU ADJ
Error = 5.0”
dH: = 0.7 mm
Accept