Messages
SPRINTER 100/100M/200/200M
55
E103
Too Dark !
Light up staff uniformly.
E104
No Staff !
Check targeting.
E105
Invalid Entry !
Check the entry / input.
E106
Out Of Level !
Level the instrument.
E107
Memory Full !
Set internal memory to OFF and continue
measurement without recording OR download
stored data to an external device and continue
measurement with internal memory ON after
deleting all the internal memorized data.
E108
Data File Error !
Data file error.
E109
Low Memory !
Prepare to download data to external device, in
order to continue further measurement with
recording after deleting all the internal memo-
rized data.
E110
Target Too Close !
Move staff or instrument further apart.
E111
Target Too far !
Move staff or instrument closer together.
E112
Too Cold !
Stop working, external temperature is outside
the instrument operating temperature.
E113
Too Hot !
Stop working, external temperature is outside
the instrument operating temperature.
No.
Error message
Counter measure / causes