68
Check & Adjust
Next step
IF the results are
THEN
to be stored
overwrites the old adjustment errors with the new
ones, if the
Use
status is set to
Yes
.
to be determined
again
rejects all new determined adjustment errors and
repeats the whole procedure. Refer to step 3. of paragraph
"Combined adjustment procedure step-by-step".
<No.of Meas:>
Shows the number of runs executed. One run consists of
a measurement in face I and face II.
<
σ
l Comp:>
and similar lines show the standard deviations of the deter-
mined adjustment errors. The standard deviations can be calculated from
the second run onwards.
It is recommended to measure at least two runs.
8.
if more runs have to be added. Continue with step 3.
OR
to accept the measurements and to proceed to
TOOLS
Adjustment Results
. No more runs can be added later.
Step
Description