The following steps
to
shall not be done in the same room as
the mounting/dismounting takes place.
4.
Clean the fine filter cartridge with fresh clean compressed air, for
example with compressed air spray. Clean it from the inside of the
fine filter cartridge to the out and not the vice-versa.
5.
Clean the coarse suction filter with fresh clean compressed air, for
example with compressed air spray. Clean it from the inside out and
not the vice-versa.
6.
Clean the coarse discharge filter gently with a small brush, to remove
dust laying down on the aluminum ribs.
☞
If some particles of dust are clearly stuck inside the mesh
fibers, do not try to remove them to not force it more
inside or to avoid damaging the mesh.
☞
Do not use compressed air to clean the discharge filter.
NOTICE
Damaged filter cartridge, filter mesh or seals
If the device is used with damaged filter cartridge, filter mesh or seals, it will
irretrievably damage inner components.
Precautions:
▶
The BLK2GO needs technical assistance from a Leica Geosystems author-
ised service centre if:
•
Any filter mesh is damaged.
•
The cartridge is deformed and cannot be properly plugged inside.
•
In case of damages to the O-rings which seal the filter to the impeller
housing.
Care and Transport
35