Aperio GT 450 DX User’s Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
57
Chapter 5: Maintenance
6.
When the filter is completely dry, slide it back into position.
7.
If you are performing the entire six-month maintenance routine, continue to the next procedure.
8.
If you are ready to close the scanner cover and restart the scanner, follow the steps in
Interior Maintenance” on page 58
Clean the Slide Racks
`
Frequency: at least every six months
.
To clean the slide racks:
1.
Check the racks for any damage or accumulated glass dust and mounting media.
2.
If a rack is damaged, replace it immediately.
3.
Use compressed air or a cleaning solvent to clean all grooves in the rack, ensuring the racks are
clean and smooth.
(We recommend Xylene cleaning solvent.)
4.
If you are performing the entire six-month maintenance routine, continue to the next procedure.
5.
If you are ready to close the scanner cover and restart the scanner, follow the steps in
Interior Maintenance” on page 58
Clean the Scanner Cover
`
Frequency: at least every six months
.
To clean the scanner cover:
1.
Wipe the outer cover of the scanner with a damp cloth.
2.
Immediately dry the cover using a dry cloth.