Aperio GT 450 DX User’s Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
58
Chapter 5: Maintenance
Clean the Touchscreen
`
Frequency: at least every six months
.
To clean the touchscreen:
1.
Spray standard monitor cleaner directly on a clean non-scratch cloth. (Avoid spraying the touchscreen directly to
ensure that no liquid gets inside the scanner.)
2.
Thoroughly wipe down the touchscreen with the cloth.
3.
If you are ready to close the scanner cover and restart the scanner, follow the steps in
Interior Maintenance” on page 58
.
Restart the Scanner After Interior Maintenance
Follow these steps to restart the scanner after you have accessed the scanner interior.
1.
Pivot the VPU into the closed position.
2.
Secure the VPU in place by rotating the VPU latch forward or backward 180 degrees, depending on its current
position.