Aperio GT 450 DX User’s Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
65
Chapter 6: Troubleshooting
6.
Close the scanner cover:
a.
Pivot the VPU into
place.
b.
Rotate the VPU latch.
c.
Slide the cover until it clicks
closed.
For detailed steps, see
“Restart the Scanner After Interior Maintenance” on page 52
.
7.
Tap
Restart Scanner
on the touchscreen interface, and wait for the scanner to complete the restart process.
8.
If the issue persists, call Leica Biosystems Technical Services.
1001: Scanner cannot initialize
`
Cause:
The scanner cannot complete the initialization process.
Follow these steps:
1.
Open the cover and access the interior:
a.
Open scanner cover.
b.
Rotate the VPU latch.
c.
Pivot out VPU
For detailed steps, see
“Open the Scanner Cover and Access the Interior Components” on page 42
2.
Check if there is a slide on the stage or partially on the stage.