Aperio GT 450 DX User’s Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
72
Chapter 6: Troubleshooting
9.
Close the scanner cover:
a.
Pivot the VPU into
place.
b.
Rotate the VPU latch.
c.
Slide the cover until it clicks
closed.
“Restart the Scanner After Interior Maintenance” on page 52
10.
Tap
Restart Scanner
on the touchscreen interface, and wait for the scanner to complete the restart process.
11.
If the issue persists, contact Leica Biosystems Technical Services.
2001: Slide handling error at rack gripper, lift, or carousel.
`
Cause:
There is an obstruction near the rack gripper, lift, or carousel.
Follow these steps until the issue is resolved:
1.
Open the cover and access the interior:
a.
Open scanner cover.
b.
Rotate the VPU latch.
c.
Pivot out VPU
“Open the Scanner Cover and Access the Interior Components” on page 42
.
2.
Take photos of the obstruction.
WARNING:
Do not attempt to recover broken slides. Call Leica Biosystems Technical Services.