Chapter 6: Troubleshooting
9000: Scanner cover is open
`
The scanner cover is not closed completely.
Follow these steps until the issue is resolved:
1.
Slide the cover to the closed position, making sure the cover aligns with rear latches and clicks closed:
Rack Warnings and Solutions
Rack warnings indicate a problem with the rack or one or more slides in the rack. The scanner can continue scanning when
there is a rack warning.
CAUTION:
If you need to remove a rack before all slides in the rack are scanned, first take note of the rack and
slide status. After you remove a rack, the rack’s scan status is no longer available on the Home screen.
1005: Cannot process rack.
`
Cause:
There is an issue with a rack that prevents scanning.