Chapter 6: Troubleshooting
Aborted Message Appears on All Slides
Follow these steps in order until the issue is resolved:
1.
Remove any completed slide racks from the carousel.
2.
Open the cover and access the interior:
a.
Open scanner cover.
b.
Rotate the VPU latch.
c.
Pivot out VPU
“Open the Scanner Cover and Access the Interior Components” on page 42
.
3.
Check if there is a slide on the stage or partially on the stage.
CAUTION:
Restarting the scanner with a slide on the stage might damage the slide.
4.
If there is a slide on the stage, carefully remove it from the stage without touching the surrounding components.
5.
Close the scanner cover:
a.
Pivot the VPU into place.
b.
Rotate the VPU latch.
c.
Slide the cover until it clicks
closed.