Aperio GT 450 DX User’s Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
46
This chapter contains a maintenance schedule and procedures for maintaining your Aperio GT 450 DX scanner.
Call Leica Biosystems Technical Services if you cannot perform a maintenance routine or if you discover a problem with your
scanner.
Before performing maintenance, we recommend you watch the maintenance videos available on the touchscreen. See
.
CAUTION:
It is important that you perform the maintenance procedures listed below on a regular schedule.
Not doing so could result in scanner misalignment which could impair image quality.
Maintenance Schedule
To keep your Aperio GT 450 DX scanner in optimal working condition, follow these maintenance routines.
Frequency
Maintenance Task
Procedure
Daily
(Recommended)
Restart the scanner.
“Restart the Scanner” on page 50
Every Six Months
1.
Clean the objective and Koehler
“Clean the Objective and Koehler” on page 51
2.
Clean the stage slide tray
“Clean the Stage Slide Tray” on page 54
3.
Clean the carousel
“Clean the Carousel” on page 55
4.
Clean the slide racks
“Clean the Slide Racks” on page 57
5.
Clean the fan filter
“Clean the Fan Filter” on page 55
6.
Clean the touchscreen
“Clean the Touchscreen” on page 58
7.
Clean the scanner cover
“Clean the Scanner Cover” on page 57
Once a year
Schedule annual maintenance visit by
Leica Biosystems Technical Services.
Call Leica Biosystems Technical Services.
Refer to
“Customer Service Contacts” on page 11
.
5
Maintenance