Controls
18
ARveo / Ref. 10 747 384 / Version 04
Interface panels with GLOW800
25
26
25 DVI In from endoscope camera (HDMI compatible)
26 XGA in from IGS systems
27
28
29
27 DVI out for external monitor (note the requirements regarding
EMC on this page)
28 SDI out for IGS systems
29 Ethernet***
*** Only systems verified by Leica Microsystems (Schweiz) AG may be
connected here.
External Monitor
Interface panel stand
Ferrite
1
Ferrite
2
Ferrite
2
Ferrite
1
Requirements regarding EMC
To reduce emission and ensure EMC compatibility of external monitors
connected to the DVI outlet (see picture).
The cable has to be equipped with 2 ferrites each (outlet at stand an inlet
at Monitor).
Specification for the ferrites:
Würth 74271622 (tested ferrite)
Würth 74271112 (tested ferrite)
Ferrites with the same frequency/impedance charateristic can also be
used.