Operation
ARveo / Ref. 10 747 384 / Version 04 39
You can read the currently set magnification on the
display (1) on the Leica M530 Optics carrier.
1
Manually adjusting the magnification (zoom)
NOTE
Destruction of the magnification motor.
X
Only adjust the magnification manually if the magnification
motor is defective.
If the magnification motor fails, the magnification can be manually
adjusted using the rotary knob (2).
X
Push in rotary button (2).
X
Set the desired magnification by turning the knob.
2
8.3.6 Setting the working distance (WD, focus)
WARNING
Danger of serious damage to tissue due to incorrect working
distance.
X
When using lasers, always set the working distance of
the microscope to laser distance and lock the microscope
in position.
X
Do not adjust the rotary button for manual setting of
the working distance while using the laser.
WARNING
Danger of injury to the eyes due to laser radiation.
X
Never point the laser directly or indirectly via reflecting
surfaces to the eyes.
X
Never point the laser to the eyes of the patient.
X
Do not look into the laser beam.
You can adjust the working distance using the footswitch/
handswitch or the "working distance" adjustment bar on the
"Main" menu screen of the control unit.
On the touch panel monitor in the "Main" menu screen
X
Press the
or
button on the bar for adjusting the
working distance.
– or –
X
Press the working distance adjustment bar directly.
The working distance changes.