Care and maintenance
ARveo / Ref. 10 747 384 / Version 04 57
X
Open access door (2) for lamp insert.
The illumination push-button (item 2, page 7) flashes orange.
4
3
1
2
CAUTION
Danger of skin burns. The lamp insert gets very hot.
X
Check that the cover has cooled before you replace the
lamp.
X
Remove the defective lamp insert (1 or 3) and install a new
lamp insert (available from Leica Microsystems).
When installing the lamp insert, make sure that the
arrow (4) is pointing to the left.
X
Close the access door again.
The illumination push-button (item 2, section 6.2) lights up
green.
Whenever you replace a bulb, reset the bulb's hour meter
to 0 by double-clicking the "Reset" button, see page 50.
11.5 Changing fuses
X
Remove the cover (1) with a screwdriver.
1
Lorem ipsum
X
Unscrew the fuse holders, remove the fuses and replace them
with the fuse type specified in the table (2) on the back side of
the cover (1).
2
F1, F2
PRIM T6.3 AH 250V Stand
F3, F4
PRIM T6.3 AH 250V Main Xenon Power Supply
F5, F6
PRIM T6.3 AH 250V Main Xenon Power Supply
F7, F8
PRIM T6.3 AH 250V Video System
F9
SEC T4 AH 250V Video System