21
3.5 Assembling the tubes:
HCI B 22
Binocular tube with 45° viewing angle
Field of view index up to 22
Eyepiece diameter 30 mm for HC PLAN 10x/20 or
22 eyepieces
Interpupillary distance setting: 55–75 mm
HCI 3T 22
Trinocular tube, 45° viewing angle
Light path:
100 % vis
50 % - 50 %
100 % photo
Field of view index up to 22
Eyepiece diameter 30 mm for HC PLAN 10x/20 or
22 eyepieces
Interpupillary distance setting: 55–75 mm
HCI BV 22
Ergo binocular tube with 15°–50° viewing angle
Field of view index up to 22
Eyepiece diameter 30 mm for HC PLAN 10x/20 or
22 eyepieces
Interpupillary distance setting: 55–75 mm
Using a hexagonal screwdriver, slacken the
clamp screw (20.1, 21.1, 22.1) on the side of the
tube change mount on the stand and remove the
black cap.
Mount the tube so that the guide pin snaps in
position and the Siedentopf binocular element
(20.3, 21.3, 22.3) points upwards as a V shape.
Retighten the clamp screw. Hold on to the tube
until the clamp screw is tightened as otherwise it
might fall off.
Insert the eyepieces into the tube ports as far as
the stops.
Insert the eyepiece with adjustable eyelens into
the right-hand port, where it is easier to handle.
Fig. 20
HCI B 22
1
Clamp screw,
2
Tube port,
3
Siedentopf binocular element
Fig. 21
HCI 3T 22
1
Clamp screw,
2
Tube port,
3
Siedentopf binocular element,
4
Photo/TV port,
5
Beamsplitter switch rod
2
3
2
3
1
1
5
4