31
Lamphousing for transmitted light illumination
(44)
for 12 V 100 W halogen lamps with single-lens
aspherical collector and heat protection filter is
an integral part of the transmitted light illuminati-
on column.
Assembling and changing the 12 V 100 W
halogen lamp
Disconnect from the power supply.
Slacken the fastening screws (44.2) and remove
the lamphousing casing (44.1).
Carefully remove the defect lamp from the pin
base.
Insert a new 12 V 100 W lamp, still in its protective
cover, into the pin base (45.1) as far as the stop.
n.b.: Leave the lamp’s protective cover on until it
is fully inserted.
Avoid making fingerprints or wipe off immediate-
ly.
Close the lamphousing again and fix in position.
Fig. 44
Lamphousing for transmitted light illumination
1
Cover,
2
Cover fixing screws
Fig. 45
Lamphousing for transmitted light, cover removed
1
Lamp holder (pin base) with 12 V 100 W halogen lamp,
2
Collec-
tor,
3
Heat protection filter
1
2
2
1
3